Dosimetri av CT-skanner. Behov for anbefalinger
|
|
- Mari Kristoffersen
- 8 år siden
- Visninger:
Transkript
1 Dosimetri av CT-skanner. Behov for anbefalinger Hans Bjerke Gardermoen,
2 CTDI 100, c Dosimetry in CT ICRU REPORT 87 CTDI 100, p 1 cm CTDI a C K 1 N T K a z dz CTDI w C K,PMMA,w CT pitch factor 1 3 d N T C K,PMMA,100,c 2 3 C K,PMMA,100,p CTDI vol CTDI w / CT pitch factor C K,PMMA, w / CT pitch factor DLP CTDI vol L
3 CT-dosimetry with pencil chamber free in air NRPA calibration set up for 2 cm slit
4 CT-calibration at the NRPA (a) The reference pencil chamber, at 100 cm from the tube focus, on a plastic holder that is fastened on a stand. It is positioned 2 cm behind the 5x5 cm 2 square aperture. (b) The monitor chamber that can be used for determination of the irradiated time can be positioned here (it is not completely seen in the picture). (c) The filtration wheel for the different radiation qualities and a lead shutter behind
5 CT-dosimetry - calibration techniques At SSDL: The radiation qualities will be the RQTs (IEC 61267:2005) Standard Radiation Quality X-ray tube voltage kv Nominal added filter for RQT mm Cu 1. HVL mm Al RQT ,2 6,9 RQT ,25 8,4 RQT ,3 10,1
6 Two approaches for CT dosimetry 100mm active length 0,3 mm active lenght CT ion chamber CT-SD16 solid state detector
7 CT-dosimetry with CTDP in phantom Exradin A3 CT-dose profiler (CTDP) eller Farmer kammer
8 RTI, Sweden: Barracuda & Piranha CT-SD16 Probe
9 Quality control program Example of a quality control program for a CT unit: Physical parameter Level of expertise Frequency Priority Remedial level Image noise A D/W H Baseline ± 20% B 1 y* H *more scans/parameters CT number values A D/W H Water: ± 10 HU Other materials ± 20 HU B 1 Y* H *more scans/parameters CT number uniformity B 1 y H > ± 1.5% CTDI A 1 y H Baseline ± 20% Irradiated slice thickness B 1 y H > ± 20% or 1 mm Image slice thickness B 1 y M > ± 20% or 1 mm High contrast resolution B 1 y H Baseline ± 20%
10 Dosimetry - CTDI in Perspex Phantoms Insert to plug holes Head phantom Body phantom (or annulus to fit over head phantom) 150 mm long
11 CT-dosimetry, spectre and HVL HVL~ between 5 and 9 mm Al
12 ICRU CT phantom and the beam From ICRU Report 87 3 x 20 cm sections, Ø=15 cm polyethylene
13 H(L) profiles CT dose as a function of phantom length
14 HVL measurements Traditional method Real-time method CT dose as a function of phantom length
15 Dosimetry - CTDI in air When commissioning a CT scanner, it is valuable to measure the CTDI in air under the following conditions: For all beam shaping filters For all nominal slice widths For all clinical kv settings For a range of scan times For a range of ma settings
16 CT-dosimetry 1. What is the x-ray installation team doing? 2. How are the medical physicist designing the acceptance test? 3. What measurements are in the control program? 4. Regulatory needs (collection of representative doses)? DLP CDTI vol How do we establish traceability for CT-dosimetry? Nordic statement from 2012: The triple A concept: Awareness, Appropriateness and Audit.
17 Protokoller for CTDI og DLP Trude sendte ut en e-post 12. september 2014 og ba om tilbakemelding om: 1. Har sykehuset protokoller/prosedyrer for måling av CTDI air og CTDI w? 2. Hvis ja, ber vi om at prosedyrene sendes elektronisk til undertegnende. 3. Brukes prosedyrene? 4. Hvis ja, ber vi også om at måleresultatene fra siste statuskontroll legges ved (tilstrekkelig med 1 som eksempel fra hver). 5. Hva slags detektor måler dere med (pennkammer, diode eller evt andre. Angi fabrikant)? 6. Ser dere spesielle utfordringer/problemer som finnes for CT-dosimetri?
18 Sendt til - svar fra Sendt til: Arne Erikson, Bengt-Erik Johansson, Per Johan Vikström Kristin Ramberg, Liv-Berit Moe, Rune Hafslund, Steinar Tveiten og Hilde Olerud Svar fra: Elisabeth U. Aasland (SUS), Arne Erikson (UNN), Anette Aarsnes (OUS), Kirsten Nygaard Bolstad (HUS), Erlend Andersen (SSK) og Kristin Ramberg.
19 Protokoller/prosedyrer for måling av CTDI air og CTDI w? HF Har Prosedyre Brukes Resultater SUS Ja Metodebok for CT mottaks- og statuskontroll Ja/RTI OUS Ja Metodebok Kvalitetskontroller CT Ja Ja UNN Ja CT 12 dose Ja Ja HUS Ja Stråleteknisk statuskontroll CT Ja Ja SSHF Ja OUS Ja Ja SOH Ja SIS Ja Ja Ja
20 Opplysningsside Mottaks- statuskontroll St Olavs hospital Excel ark fra SIS Danmark Kontrolltype: Statuskontrol Sykehus: St. Olavs Hospital Avdeling: AHL21CT Romnummer: 341.U1.036 Skanner-fabrikat: Siemens Skanner-type: Somatom Sensation 64 Registreringsnummer Merida nr: 685 Installasjonsdato: CT-Skanner AHL21CT Antall mulige kv 4 Antall fysiske filtreringer 2 (body/head) Måleutstyr til kv-målinger Raysafe XI kalibreringsdato kalibreringsdato Måleprobe til dosis-målinger Raysafe CT-probe kalibreringsdato kalibreringsdato Fantom til dosis-målinger Fluke CTDI Fantom til billedkvalitet Catphan 600 Software til billedkvalitetsanalyse CTQALite Dato for kontroll Kontroll utført av KR SEJ COW telefonnummer adresse kristin.ramberg@stolav.no
21 Opplysningsside Mottaks- statuskontroll 0 Sygehus Utført av: KR SEJ COW St. Olavs Hospital Utført dato: 31. juli 2014 Fabrikat : Siemens Avdeling: AHL21CT Rum nr.: 341.U1.036 Reg.nr.: Merida nr: 685 Test: Dose CTDI w St. Olavs Hospital HF Statuskontrol Version: 1.0 Revideret den: Rørspænding [kv] ,8 24x1, ,8 24x1, x0, x0,6 Rørspænding [kv] CTDI w - BODY-filter Kollimering [mm] CTDI w - HEAD-filter Kollimering [mm] Rørstrøm [ma] Rotationstid [s] mas Rørstrøm [ma] Rotationstid [s] x0, x0, x0, Rørspænding [kv] Evaluering Målt værdi mod skannerværdi -5,6% -4,9% -5,6% Kalibreringsfaktor Ionkammernormering Detektorkonfiguration Kalibreringsfaktor Detektorkonfiguration Ionkammernormering CTDI w - Siemens anbefalte innstillinger: Body + Head mas CTDI vol -målinger [mgy] Center Perifer kl. 12 Perifer kl. 15 2,568 4,707 4,555 4,492 4,52 1,451 2,697 2,7 2,568 2,684 1,668 3,135 3,016 2,974 2,993 0,96 1,786 1,796 1,708 1,781 Center Perifer kl. 15 Perifer kl. 18 CTDI vol -målinger [mgy] Perifer kl. 12 Perifer kl. 18 Perifer kl. 21 Perifer kl. 21 1,341 1,412 1,433 1,431 1,371 2,823 2,916 2,949 2,945 2,836 9,291 9,591 9,646 9,678 9,34 Modtagekontrol Beregnede værdier CTDI w [mgy] CTDI w /mas [mgy] 13,5 0,065 7,8 0,037 14,3 0,068 8,3 0,040 Modtagekontrol Beregnede værdier CTDI w [mgy] CTDI w /mas [mgy] 7,7 0,041 16,0 0,084 52,6 0,138 Modtagekontrol Beregnede værdier Aflæst værdi CTDI vol (skanner) [mgy] 14,15 8,23 15,14 8,8 Aflæst værdi CTDI vol (skanner) [mgy] 8,17 16,83 55,75 Evaluering Målt værdi mod skannerværdi -4,3% -4,7% -5,5% -5,4% CTDI vol -målinger [mgy] Aflæst værdi Evaluering CTDI vol Målt værdi Kalibreringsfaktonormering [mm] konfiguration kl. 12 kl. 15 kl. 18 kl. 21 [mgy] [mgy] [mgy] værdi Ionkammer- Kollimering Detektor- Perifer Perifer Perifer CTDI CTDIw /mas Rørstrøm [ma] Rotationstid [s] mas Center Perifer w (skanner) mod skanner ,8 24x1, , ,68 5,045 4,781 4,565 4,724 14,2 0,064 14,83-4,5% ,8 24x1, ,95 12,38 12,49 12,5 12,08 42,4 0,129 45,41-6,5%
22 Opplysningsside Mottaks- statuskontroll St. Olavs Hospital HF Statuskontrol 0 Sygehus Utført av: KR SEJ COW St. Olavs Hospital Utført dato: 31. juli 2014 Fabrikat : Siemens Avdeling: AHL21CT Rum nr.: 341.U1.036 Reg.nr.: Merida nr: 685 Test: Dose CTDI w Version: 1.0 Revideret den: Statuskontrol Modtagekontrol CTDI w - BODY-filter Oplysninger CTDI vol -målinger CTDI [mgy] vol -målinger [mgy] Beregnede værdier Beregnede Aflæst værdier værdi Aflæst værdi Evaluering Evaluering CTDI vol Perifer Perifer CTDI CTDI w /mas Center CTDI Målt værdi Målt værdi Rørspænding Kalibreringsfaktor [mm] konfiguration mas CTDI w CTDI w /mas vol Ionkammernormering kl. 18 kl. 21 Kollimering Detektor- Rørstrøm [ma] Rotationstid [s] w (skanner) skannerværdi [kv] Kalibreringsfaktor normering kl. 12 kl. 15 kl. 18 kl. 21 [mgy] [mgy] [mgy] Ionkammer Perifer Perifer Perifer Center Perifer (skanner) mod skannerværdi [mgy] kontrol mod modtage- [mgy] ,8 24x1, ,532104,676 4,574 2,568 4,456 4,707 4,439 4,555 4,492 13,4 4,52 0,064 13,5 14,15 0,065-5,1% 14,15-0,9% -4,3% ,8 24x1, , ,702 2,752 1,451 2,688 2,697 2,666 2,7 2,568 7,92,684 0,038 7,8 0,037 8,23-3,5% 8,23-4,7% 1,2% x0, , ,007 3,047 1,668 2,884 3,135 3,016 2,96 2,974 14,12,993 0,067 14,3 15,14 0,068-6,6% 15,14-1,2% -5,5% x0, , ,874 1,828 0,96 1,724 1,786 1,783 1,796 1,708,5 1,781 0,040 8,3 0,040 8,8-3,8% 8,8-5,4% 1,7% Statuskontrol Modtagekontrol CTDI w - HEAD-filter Oplysninger CTDI vol -målinger CTDI [mgy] vol -målinger [mgy] Beregnede værdier Beregnede Aflæst værdier værdi Aflæst værdi Evaluering Evaluering CTDI CTDI vol Rørspænding Kalibreringsfaktonormering [mm] konfiguration faktor normering kl. 12 kl. 15 kl kl kl. [mgy] 18 kl. 21 [mgy] [mgy] [mgy] værdi [mgy] kontrol værdi Ionkammer- Kollimering Detektor- Perifer Perifer CTDI w CTDI vol Målt værdi Målt værdi Kalibrerings- Ionkammer CTDI CTDI w /mas Rørstrøm [ma] Rotationstid [s] mas Center w /mas Center Perifer Perifer Perifer Perifer w (skanner) mod (skanner) skanner- mod modtage- skanner- [kv] x0, Utgår 1,341 på status 1,412 1,433 1,431-1,371 7,7-0,041 8, ,6% x0, , ,957 2,981 2,823 2,916 2,98 2,883 2,949 2,945 16,22,836 0,086 16,0 16,83 0,084-3,5% 16,83-4,9% 1,5% x0, , ,572 9,598 9,291 9,687 9,591 9,364 9,646 9,678 52,7 9,34 0,139 52,6 55,75 0,138-5,4% 55,75-5,6% 0,2% Statuskontrol Modtagekontrol CTDI w - Siemens anbefalte innstillinger: Body + Head Oplysninger CTDI vol -målinger CTDI [mgy] vol -målinger [mgy] Beregnede værdier Beregnede Aflæst værdier værdi Aflæst værdi Evaluering Evaluering CTDI vol Rørspænding Kalibreringsfaktonormering [mm] konfiguration faktor normering kl. 12 kl. 15 kl kl kl. [mgy] 18 kl. 21 [mgy] [mgy] [mgy] værdi [mgy] kontrol værdi Ionkammer- Kollimering Detektor- Perifer Perifer CTDI w CTDIw /mas Rørstrøm [ma] Rotationstid [s] mas Center Målt værdi Målt værdi Ionkammer Perifer Perifer Perifer Perifer CTDI Center w CTDI w /mas vol Kalibrerings- (skanner) mod (skanner) skanner- mod modtage- skanner- [kv] ,8 24x1, ,5 2, ,026 4,814 2,68 4,812 5,045 4,675 4,781 4,565 14,34,724 0,065 14,2 14,83 0,064-3,9% 14,83-4,5% 0,7% ,8 24x1, Utgår 11,95 på status 12,38 12,49 12,5-12,08 42,4-0,129 45, ,5% -
23 Dosimetriutstyr, type, målenheter Det var ellers store mangler i prosedyrene med hensyn til angivelse av: Fabrikant, type utstyr og serienummer. Noe mer info fra kommentarer. Manglende enheter på dosimetristørrelser både i prosedyre og rapporter. Usikkerhet i målingene var ikke angitt.
24 Detektor problemer i CT dosimetri HF Detektor Problem SUS Piranha og CT Dose Profiler Ikke tilgang på «Service på CT» OUS RTI Barracuda + pennkammer Unfors 10 cm + utlesningsenhet Kobling av CTDI og SSDE Doseberegn. med autom. dosemod UNN Unfors Xi External Detector CT Ikke helikalt skann ikke standardiserte målinger HUS Unfors Xi External Detector CT Måler med sekvensielle snitt, mens det brukes helikale protokoller SSHF SOH Pennkammer RaySafe (XI CT detector) Raysafe (tidligere Unfors), med tilhørende CT-probe. Estimering av effektivdose, pasientstørrelse, AEC, helkale opptak Manualer er ikke forklarende for beregnet stråledose.
25 Eksempel på usikkerhet: Dose til foster etter CT-undersøkelse Antar at du beregner forsterdose til 70 mgy. Videre antas følgende: Du har ikke kalibrert kammer Du har målt opp med smalt fantom (L=15 cm) Du godtar CTDI ± 20% Rektangelfordelt ± 20% vil gi usikkerhet 20%/ 3 = 11,5 % ± 10% (k=1) + 20% ± 11,5% (k=1) Sum ± 25% (k=1) Hva er usikkerheten i målingen din? ± 50% (95 % konfidensintervall, k=2) Fosterdose er (70 ± 35) mgy
26 Dose til foster etter CT-undersøkelse Antar at du beregner forsterdose til 70 mgy. Videre antas følgende: Du har kalibrert kammer Du har målt opp med langt fantom Du godtar CTDI ± 20% ± 2% (k=1) ± 1% (k=1) ± 11,5% (k=1) Sum ± 12% (k=1) Hva er usikkerheten i målingen din? ± 24% (95 % konfidensintervall, k=2) Fosterdose er (70 ± 17) mgy
27 Anbefalinger 1. Kjøp ICRU report 87 Radiation Dose and Image-Quality Assessment in Computed Tomography, Last ned SIS: Protokol for modtage- og statuskontrol af CTskannere, Sundhedsstyrelsen, Vurder å bruke Excelark fra SIS laget av Asbjørn Seegert. (Anbefalt av St Olavs) Hvis dere har programmer som følger med detektor be om å få eksport-fil som kan importeres i ark fra SIS. 4. Kalibrer CT-kammer jevnlig (vi tilbyr dette). 5. I tillegg anbefales å besøke hverandre for å sammenligne måleutstyr og målemetoder. Arranger dosimetriseminar med praktisk CT-måling.
28 Buy this and read it!!
Statuskontroll av Leverandør X, 16 snitts CT skanner, Sykehus X
Statuskontroll av Leverandør X, 16 snitts CT skanner, Sykehus X Seksjon for diagnostisk fysikk Intervensjonssenteret Oslo universitetssykehus 2 Rapport statuskontroll Laboratorium: X Sykehus: X Dato: 27.7.2011
DetaljerReferansedosimetri for protonterapi
Referansedosimetri for protonterapi Hans Bjerke Værnes, 03.03.2014 Oversikt 1. Helsepolitikk og vitenskap 2. Norsk satsing i kjernefysikk historisk perspektiv 3. Definisjoner, z ref, R p, SOBP, R res,
DetaljerProtokoll for kontroller av digitale mammografisystemer
Protokoll for kontroller av digitale mammografisystemer Mål Samle ei arbeidsgruppe for å komme fram til en nasjonal protokoll for kontroller av digitale mammografisystemer Mammografiscreening og Strålevernet
DetaljerKvalitetskontroll ved UUS. Charlotte Kile Larsen Kompetansesenter for Diagnostisk Fysikk Ullevål Universitetssykehus HF
Kvalitetskontroll ved UUS Charlotte Kile Larsen Kompetansesenter for Diagnostisk Fysikk Ullevål Universitetssykehus HF KDF - røntgen 5,5 stillinger 18 sykehus 250 modaliteter sjekkes hvert år CT Gjennomlysning
DetaljerPSi Apollo. Technical Presentation
PSi Apollo Spreader Control & Mapping System Technical Presentation Part 1 System Architecture PSi Apollo System Architecture PSi Customer label On/Off switch Integral SD card reader/writer MENU key Typical
DetaljerIndikasjoner. Generelt. CT nevroendokrin tumor. Parameter Teknikk Kommentar. Nevroendokrin tumor.
Prosedyre Gyldig fra: 09.03.2018 Organisatorisk plassering: HVRHF - Helse Vest RHF regionale dokumenter Dok. eier: Aslak Bjarne Aslaksen Dok. ansvarlig: Mowinckel-Nilsen, Mia Louise Indikasjoner Nevroendokrin
DetaljerEn enkel database over utstyr og målinger ved HUS. Helge Pettersen, Helse Bergen. 04.11.2013 Helge Pettersen 1
En enkel database over utstyr og målinger ved HUS Helge Pettersen, Helse Bergen 04..203 Helge Pettersen Introduksjon Spørsmål : Hvordan skal man holde oversikt over modaliteter? I fem-seks forskjellige
DetaljerStråledoser ved undersøkelser av abdomen Nasjonalt strålevernsmøte 5. november 2013
Stråledoser ved undersøkelser av abdomen Nasjonalt strålevernsmøte 5. november 13 Anita Fosterud Reitan Strålevernkontakt BDA Camilla Lunder Jensen Strålevernkoordinator Stråledoser ved undersøkelser av
DetaljerKROPPEN LEDER STRØM. Sett en finger på hvert av kontaktpunktene på modellen. Da får du et lydsignal.
KROPPEN LEDER STRØM Sett en finger på hvert av kontaktpunktene på modellen. Da får du et lydsignal. Hva forteller dette signalet? Gå flere sammen. Ta hverandre i hendene, og la de to ytterste personene
DetaljerGenerell utredning og kontroll av abdominale tilstander der mer organspesifikk prosedyre ikke er indisert.
Prosedyre Gyldig fra: 09.03.2018 Organisatorisk plassering: HVRHF - Helse Vest RHF regionale dokumenter Dok. eier: Aslak Bjarne Aslaksen Dok. ansvarlig: Mowinckel-Nilsen, Mia Louise Indikasjoner Generell
DetaljerTB-615 / TB-617 Wireless slim keyboard. EN User guide SE Användarhandledning FI Käyttöohje DK Brugervejledning NO Bruksanvisning
TB-615 / TB-617 Wireless slim keyboard EN User guide SE Användarhandledning FI Käyttöohje DK Brugervejledning NO Bruksanvisning EN User guide You have bought a wireless keyboard to use with Windows XP,
DetaljerKvalitetskontroll røntgen Dose til homogent fantom. Jacob Nøtthellen 18.11.2008
Kvalitetskontroll røntgen Dose til homogent fantom Jacob Nøtthellen 18.11.2008 A.Bakgrunn B.Metodikk C.Resultater D.Spin-off A.Bakgrunn Spørsmål fra radiolog etter kontroll av røntgenlab rundt 1990: Er
DetaljerMøtearena for diagnostisk fysikk
Møtearena for diagnostisk fysikk Strålevernets Novembermøte 4/11-2013 Radisson Blu Hotell, Gardermoen National reference group: Health region North Samuel Kuttner Arne Erikson Health region Middle Kristin
DetaljerIMR Erfaringer med en ny iterativ rekonstruksjonsmetode for CT
IMR Erfaringer med en ny iterativ rekonstruksjonsmetode for CT MedFys 2016 Siri Fløgstad Svensson Fysiker, røntgen og CT, Intervensjonssenteret E-post: sirifl@ous-hf.no sirifl@so-hf.no Ulike typer iterativ
DetaljerOm strålevernet Om den nye forskriften. Om dosestørrelser knyttet til ulike modaliteter Hvem tar utfordringen?
Ny strålevernslovgivning og behov for IT løsninger der RiS/PACS leverandørene vil være en sentral aktør. Hilde M. Olerud, dr.ing seksjonssjef, Statens Strålevern Om strålevernet Om den nye forskriften
DetaljerCT doser i stråleterapien. Noe å bekymre seg for?
CT doser i stråleterapien Noe å bekymre seg for? Diagnostikkfysikere Wibeke Nordhøy og Ellinor Bilet Strålefysiker Maria Sastre Stråleterapeut med videreutdanning i CT Lise Mette Sandli Strålevernsmøte
DetaljerDental Cone Beam CT. Gerald Torgersen. Kvalitetskontrollarbeid (QA) - optimalisering og dosereduksjon
Gerald Torgersen Dental Cone Beam CT Kvalitetskontrollarbeid (QA) - hvilke parametere måler vi - hvorfor måler vi de og hva brukes resultatene til? - optimalisering og dosereduksjon - litt om doser Parametere
DetaljerMarmi Plus One. Sleek, Essential, Stunning
MARMI PLUS ONE Marmi Plus One Sleek, Essential, Stunning Marmi Plus One Bianco - DME01 90x90 cm Marmi Plus One Bianco - DME01 60x120 cm Marmi Plus One Bianco MATTE Marmi Plus One Bianco - DME01 60x120
DetaljerSalting of dry-cured ham
Salting of dry-cured ham Controlled by CT scanning Torunn Thauland Håseth Skjervold-symposium 2005 Hamar, June 3rd. - en del av Matalliansen NLH - Matforsk - Akvaforsk To give you a different aspect of
DetaljerCompello Fakturagodkjenning Versjon 10 Software as a service. Tilgang til ny modulen Regnskapsføring
Compello Fakturagodkjenning Versjon 10 Software as a service Tilgang til ny modulen Regnskapsføring Dokumentopplysninger 2018 Compello AS. Med enerett. Microsoft, MS-DOS og Windows er registrerte varemerker
DetaljerHONSEL process monitoring
6 DMSD has stood for process monitoring in fastening technology for more than 25 years. HONSEL re- rivet processing back in 990. DMSD 2G has been continuously improved and optimised since this time. All
DetaljerCompello Fakturagodkjenning Versjon 10.5 As a Service. Tilgang til Compello Desktop - Regnskapsføring og Dokument import
Compello Fakturagodkjenning Versjon 10.5 As a Service Tilgang til Compello Desktop - Regnskapsføring og Dokument import Dokumentopplysninger 2018 Compello AS. Med enerett. Microsoft, MS-DOS og Windows
DetaljerSlope-Intercept Formula
LESSON 7 Slope Intercept Formula LESSON 7 Slope-Intercept Formula Here are two new words that describe lines slope and intercept. The slope is given by m (a mountain has slope and starts with m), and intercept
DetaljerExercise 1: Phase Splitter DC Operation
Exercise 1: DC Operation When you have completed this exercise, you will be able to measure dc operating voltages and currents by using a typical transistor phase splitter circuit. You will verify your
DetaljerAutomatisk dosemodulering CT
Bakgrunn Automatisk dosemodulering CT Kirsten Nygaard Bolstad Medisinsk fysiker, Helse Bergen Toshiba: Sentrering av pasient er viktig Feil sentrering -> utvidet SFOV -> økt dose Toshiba: Ved automatikk:
Detaljer5 E Lesson: Solving Monohybrid Punnett Squares with Coding
5 E Lesson: Solving Monohybrid Punnett Squares with Coding Genetics Fill in the Brown colour Blank Options Hair texture A field of biology that studies heredity, or the passing of traits from parents to
DetaljerRF Power Capacitors Class kV Discs with Moisture Protection
RF Power Capacitors Class 0-20kV Discs with Moisture Protection T H E C E R A M I C E X P E R T S RF Power Capacitors Class 0-20kV Discs with Moisture Protection The CeramTec Group is a world leader in
DetaljerMomenter. Dosimetri og bildekvalitetsparametre. Støy
Momenter Dosimetri og bildekvalitetsparametre Medisinsk fysiker Hilde Kjernlie Sæther Kompetansesenter for diagnostisk fysikk Støy og lavkontrastoppløsning Høykontrastoppløsning, Z-sensitivitet Parametre
DetaljerUnit Relational Algebra 1 1. Relational Algebra 1. Unit 3.3
Relational Algebra 1 Unit 3.3 Unit 3.3 - Relational Algebra 1 1 Relational Algebra Relational Algebra is : the formal description of how a relational database operates the mathematics which underpin SQL
DetaljerRF Power Capacitors Class1. 5kV Discs
RF Power Capacitors Class 5kV Discs T H E C E R A M C E X P E R T S RF Power Capacitors Class 5kV Discs The CeramTec Group is a world leader in the design and manufacture of complex electronic ceramic
DetaljerAnalyse av variasjon i representative doser ved Computed-tomografi undersøkelser
Analyse av variasjon i representative doser ved Computed-tomografi undersøkelser 17. November 2008 Reidun Dahl Silkoset Økte og varierende stråledose fra CT Utvikling i gjennomsnittelig dose per innbygger
DetaljerImplementeringen av ROP retningslinjen; er GAP analyser et
Implementeringen av ROP retningslinjen; er GAP analyser et effek/vt redskap? Lars Lien, leder Nasjonal kompetansetjeneste for sam
DetaljerProduktspesifikasjon og kontroll av bygningsskanning. Ivar Oveland 11. Februar 2019
Produktspesifikasjon og kontroll av bygningsskanning Ivar Oveland 11. Februar 2019 Produktspesifikasjon Norges miljø- og biovitenskapelige universitet 1 Introdusjon Arbeidet har vært 2 delt: 1. Produktspesifikasjon
DetaljerKalibrering. Hvordan sikrer Norsonic sporbarhet av måleresultatene. Ole-Herman Bjor
Kalibrering Hvordan sikrer Norsonic sporbarhet av måleresultatene Ole-Herman Bjor Norsonic kalibreringslaboratorium Norsonic kalibreringslaboratorium etablert i 1993. Akkreditert av Norsk akkreditering
DetaljerInformation search for the research protocol in IIC/IID
Information search for the research protocol in IIC/IID 1 Medical Library, 2013 Library services for students working with the research protocol and thesis (hovedoppgaven) Open library courses: http://www.ntnu.no/ub/fagside/medisin/medbiblkurs
DetaljerInteraction between GPs and hospitals: The effect of cooperation initiatives on GPs satisfaction
Interaction between GPs and hospitals: The effect of cooperation initiatives on GPs satisfaction Ass Professor Lars Erik Kjekshus and Post doc Trond Tjerbo Department of Health Management and Health Economics
DetaljerDAP-kalibrering ved sykehus
Per Otto Hetland SSDL, Statens Strålevern Gardermoen, 17.11. 2008 Innhold Motivasjon for å utføre -kalibering ved sykehusene Mulige feilkilder ved -kalibrering Anbefalinger for -kalibrering Kalibrering
DetaljerMÅLING OG VURDERING AV TEKSTUR I VEGOVERFLATER OG KOPLING TIL STØY
MÅLING OG VURDERING AV TEKSTUR I VEGOVERFLATER OG KOPLING TIL STØY Av Svein Å. Storeheier Miljøvennlige vegdekker, 2006-09-14 1 TEXTURE what can be measured The road surface is scanned by a laser system,
DetaljerNeural Network. Sensors Sorter
CSC 302 1.5 Neural Networks Simple Neural Nets for Pattern Recognition 1 Apple-Banana Sorter Neural Network Sensors Sorter Apples Bananas 2 Prototype Vectors Measurement vector p = [shape, texture, weight]
DetaljerEdge Of Dock Leveler. Size: Widths 66, 72, 78, 84. Lengths 15, 17 Lips. Capacities: 20,000 25,000 30,000 35,000 NEVERLIFT NL
Edge Of Dock Leveler Size: Widths 66, 72, 78, 84 Lengths 15, 17 Lips Capacities: 20,000 25,000 30,000 35,000 NEVERLIFT NL-7220-15 Who Uses EOD Levelers LTL trucking Companies with dedicated Truck Fleets
DetaljerRF Power Capacitors Class , 20 & 30 mm Barrel Transmitting Types
RF Power Capacitors Class 2.7, 20 & 30 mm Barrel Transmitting Types T H E C E R A M I C E X P E R T S RF Power Capacitors Class 2.7, 20 & 30 mm Barrel Transmitting Types The CeramTec Group is a world leader
DetaljerHØGSKOLEN I NARVIK - SIVILINGENIØRUTDANNINGEN
HØGSKOLEN I NARVIK - SIVILINGENIØRUTDANNINGEN EKSAMEN I FAGET STE 6243 MODERNE MATERIALER KLASSE: 5ID DATO: 7 Oktober 2005 TID: 900-200, 3 timer ANTALL SIDER: 7 (inklusiv Appendix: tabell og formler) TILLATTE
DetaljerAndrew Gendreau, Olga Rosenbaum, Anthony Taylor, Kenneth Wong, Karl Dusen
Andrew Gendreau, Olga Rosenbaum, Anthony Taylor, Kenneth Wong, Karl Dusen The Process Goal Definition Data Collection Data Preprocessing EDA Choice of Variables Choice of Method(s) Performance Evaluation
DetaljerUNIVERSITETET I OSLO ØKONOMISK INSTITUTT
UNIVERSITETET I OSLO ØKONOMISK INSTITUTT Eksamen i: ECON2915 Vekst og næringsstruktur Exam: ECON2915 - Growth and business structure Eksamensdag: Fredag 2. desember 2005 Sensur kunngjøres: 20. desember
DetaljerPARABOLSPEIL. Still deg bak krysset
PARABOLSPEIL Stå foran krysset på gulvet og se inn i parabolen. Hvordan ser du ut? Still deg bak krysset på gulvet. Hva skjer? Hva skjer når du stiller deg på krysset? Still deg bak krysset Det krumme
Detaljerstjerneponcho for voksne star poncho for grown ups
stjerneponcho for voksne star poncho for grown ups www.pickles.no / shop.pickles.no NORSK Størrelser XS (S) M (L) Garn Pickles Pure Alpaca 300 (350) 400 (400) g hovedfarge 100 (100) 150 (150) g hver av
DetaljerSoftware applications developed for the maritime service at the Danish Meteorological Institute
Software applications developed for the maritime service at the Danish Meteorological Institute Anne Marie Munk Jørgensen (ammj@dmi.dk), Ove Kjær, Knud E. Christensen & Morten L. Mortensen Danish Meteorological
DetaljerSIS Rapport 1979:5 RAPPORT NORDISK KONTAKTMØTE I DOSEMETRI. ØSTERÅS(ved Oslo) 6.og 7.september 1979. J.Flatby, H.Fosmark, H.Bjerke
STATENS INSTITUTT FOR STRÅLEHYGIENE SIS Rapport 1979:5 RAPPORT NORDISK KONTAKTMØTE I DOSEMETRI ØSTERÅS(ved Oslo) 6.og 7.september 1979 av J.Flatby, H.Fosmark, H.Bjerke State Institute of Radiation Hygiene
DetaljerMetodisk kvalitetsvurdering av systematisk oversikt. Rigmor C Berg Kurs H, mars 2019
Metodisk kvalitetsvurdering av systematisk oversikt Rigmor C Berg Kurs H, mars 2019 Oppsummering av forskning har lang tradisjon 12th century: knowledge syntheses in field of philosophy 17th century: statistical
DetaljerStatens vegvesen. Testene tilfredsstiller kravene til styrkeklasse T3 med en målt arbeidsbredden på 0,7m.
Statens vegvesen DELTA BLOC International GmbH Industriestrasse 28 2601 Sollenau Østerrike Paul Bittner Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet
DetaljerMedisinsk statistikk, KLH3004 Dmf, NTNU 2009. Styrke- og utvalgsberegning
Styrke- og utvalgsberegning Geir Jacobsen, ISM Sample size and Power calculations The essential question in any trial/analysis: How many patients/persons/observations do I need? Sample size (an example)
DetaljerMerak Un-glazed Porcelain Wall and Floor Tiles
6 Colours 3 es Absolute Collection 7 s Merak Un-glazed Porcelain Wall and Floor Tiles Merak includes seven sizes perfectly designed for seamless installation. 600 x 50mm 600 x 100mm 600 x 300mm 600 x 150mm
Detaljer5 Gy i huddose gjør det noe da? Steinar Tveiten Sentral strålevernkoordinator / Medisinsk fysiker Sørlandet sykehus HF
5 Gy i huddose gjør det noe da? Steinar Tveiten Sentral strålevernkoordinator / Medisinsk fysiker Sørlandet sykehus HF Registrering og oppfølging av høye pasientdoser ved intervensjonsprosedyrer Prosedyren
DetaljerPerpetuum (im)mobile
Perpetuum (im)mobile Sett hjulet i bevegelse og se hva som skjer! Hva tror du er hensikten med armene som slår ut når hjulet snurrer mot høyre? Hva tror du ordet Perpetuum mobile betyr? Modell 170, Rev.
DetaljerRom-Linker Software User s Manual
Rom-Linker Software User s Manual Tel.: +886-2-2274-1347 Fax. +886-2-2273-3014 Http://www.hmt.com.tw e-mail: hmtsales@hmt.com.tw Contents Catalogue Page (1) Rom-Linker Starting Screen... 01 (2) Rom-Linker
DetaljerTrust in the Personal Data Economy. Nina Chung Mathiesen Digital Consulting
Trust in the Personal Data Economy Nina Chung Mathiesen Digital Consulting Why does trust matter? 97% of Europeans would be happy for their personal data to be used to inform, make recommendations or add
Detaljerbuildingsmart Norge seminar Gardermoen 2. september 2010 IFD sett i sammenheng med BIM og varedata
buildingsmart Norge seminar Gardermoen 2. september 2010 IFD sett i sammenheng med BIM og varedata IFD International Framework for Dictionaries Hvordan bygges en BIM? Hva kan hentes ut av BIM? Hvordan
DetaljerCAMES. Technical. Skills. Overskrift 27pt i to eller flere linjer teksten vokser opad. Brødtekst 22pt skrives her. Andet niveau.
CAMES Overskrift 27pt i to eller flere linjer Technical Skills Leizl Joy Nayahangan, RN, MHCM Leizl.joy.nayahangan@regionh.dk IMPORTANCE Challenges Brødtekst 22pt of patient skrives her care Increasing
DetaljerData Sheet for Joysticks
Available with Potentiometers or Hall sensors Several handle options Small size at low installation depth The 812 series is available with several different handle options. These small joysticks are recommended
DetaljerREMOVE CONTENTS FROM BOX. VERIFY ALL PARTS ARE PRESENT READ INSTRUCTIONS CAREFULLY BEFORE STARTING INSTALLATION
2011-2014 FORD EXPLORER PARTS LIST Qty Part Description Qty Part Description 1 Bull Bar 2 12mm x 35mm Bolt Plates 1 Passenger/Right Mounting Bracket 2 12mm Nut Plate 1 Driver/Left Mounting Bracket 2 12mm
DetaljerSubstiTUBE T5 High Output
SubstiTUBE T5 High Output LED rør til elektronisk høyfrekvens forkoblingsutstyr Bruksområder _ Belysning av produksjonsområder _ Supermarkeder og varemagasiner _ Offentlige bygninger, resepsjonsområder,
DetaljerKundetilfredshetsundersøkelse FHI/SMAP
Kundetilfredshetsundersøkelse FHI/SMAP Sluttrapport pr. 20. April 2010 Alle 9 kunder av FHI s produksjonsavdeling for biofarmasøytiske produkter (SMAP) i perioden 2008-2009 mottok i januar 2010 vårt spørreskjema
DetaljerTUNNEL LIGHTING. LED Lighting Technology
TUNNEL LIGHTING TunLite Linear Designed LED Tunnel & Underpass Light The TunLite is an LED linear luminaire providing reliable solutions to cover the lighting requirements of tunnels and underpasses. It
DetaljerUNIVERSITETET I OSLO ØKONOMISK INSTITUTT
UNIVERSITETET I OSLO ØKONOMISK INSTITUTT Eksamen i: ECON1910 Poverty and distribution in developing countries Exam: ECON1910 Poverty and distribution in developing countries Eksamensdag: 1. juni 2011 Sensur
DetaljerTriCOM XL / L. Energy. Endurance. Performance.
TriCOM XL / L Energy. Endurance. Performance. L and XL - the new generation Sample charging station with chargers TriCOM L / XL Innovative charging technology The new TriCOM L - XL chargers are controlled
DetaljerThe regulation requires that everyone at NTNU shall have fire drills and fire prevention courses.
1 The law The regulation requires that everyone at NTNU shall have fire drills and fire prevention courses. 2. 3 Make your self familiar with: Evacuation routes Manual fire alarms Location of fire extinguishers
DetaljerKvalitetskontroll MR
Kvalitetskontroll MR Tone Elise Døli Orheim Seksjon for diagnostisk fysikk Disposisjon MR-systemet Kvifor kvalitetskontroll? Ulike typer kvalitetskontroll Våre testar så langt Vegen vidare MR-systemet
DetaljerINSTALLATION GUIDE FTR Cargo Rack Regular Ford Transit 130" Wheelbase ( Aluminum )
INSTALLATION GUIDE 1505-FTR Cargo Rack Regular Ford Transit 130" Wheelbase ( Aluminum ) QUICK START GUIDE Phase 1 - Assembly q 1.1 Setup... q 1.2 Cargo Rack Assembly... 3-4 5-6 Phase 2 - Installation q
DetaljerOppgave 1a Definer følgende begreper: Nøkkel, supernøkkel og funksjonell avhengighet.
TDT445 Øving 4 Oppgave a Definer følgende begreper: Nøkkel, supernøkkel og funksjonell avhengighet. Nøkkel: Supernøkkel: Funksjonell avhengighet: Data i en database som kan unikt identifisere (et sett
DetaljerDen europeiske byggenæringen blir digital. hva skjer i Europa? Steen Sunesen Oslo,
Den europeiske byggenæringen blir digital hva skjer i Europa? Steen Sunesen Oslo, 30.04.2019 Agenda 1. 2. CEN-veileder til ISO 19650 del 1 og 2 3. EFCA Guide Oppdragsgivers krav til BIMleveranser og prosess.
DetaljerProsjektet Digital kontaktinformasjon og fullmakter for virksomheter Digital contact information and mandates for entities
Prosjektet Digital kontaktinformasjon og fullmakter for virksomheter Digital contact information and mandates for entities Nordisk Adressemøte / Nordic Address Forum, Stockholm 9-10 May 2017 Elin Strandheim,
DetaljerStatens vegvesen. Rekkverket med føringsskinne av type A er i utgangspunktet samme egenskaper som det som var fullskalatestet, type B.
Statens vegvesen Stalprodukt S.A ul. Wygoda 69, 32-700 Bochnia Polen Dawid Barnaś Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet Matteo Pezzucchi /
DetaljerUNIVERSITETET I OSLO ØKONOMISK INSTITUTT
1 UNIVERSITETET I OSLO ØKONOMISK INSTITUTT BOKMÅL Utsatt eksamen i: ECON2915 Vekst og næringsstruktur Eksamensdag: 07.12.2012 Tid for eksamen: kl. 09:00-12:00 Oppgavesettet er på 5 sider Tillatte hjelpemidler:
DetaljerEn oppsummering med enkelte highlights, men solformørkelse på badet. Jacob Nøtthellen 2010
En oppsummering med enkelte highlights, men solformørkelse på badet. Jacob Nøtthellen 2010 Presentasjon av NTH-studenten Jacob Hatet ioniserende stråling Elsket biokjemi Drømte om jobb ved samlebåndet
DetaljerStatens vegvesen. Rekkverket med føringsskinne av type A er i utgangspunktet samme egenskaper som det som var fullskalatestet, type B.
Statens vegvesen Stalprodukt S.A ul. Wygoda 69, 32-700 Bochnia Polen Dawid Barnaś Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet Matteo Pezzucchi /
DetaljerNFTR Protokoll 2: CT lunger / HRCT Thorax. Om indikasjoner for protokollen. Generelt Parameter Teknikk Kommentar
NFTR Protokoll 2: CT lunger / HRCT Thorax Om indikasjoner for protokollen Protokollen innebærer primært en lungeundersøkelse, dosen er lavere enn ved standard CT thorax, og gir dårligere fremstilling av
DetaljerINNOVASJONSTOGET GÅR. - hvor er legene?
INNOVASJONSTOGET GÅR - hvor er legene? Utfordring Hvor står de medisinske fagfolkene når innovasjonen pågår, i midten eller på siden, og hva skjer fremover? Jon Endringsmotvilje? Endringsmotvilje? Helse
DetaljerStatens vegvesen. Godkjenning av StalPro Rail su vegrekkverk. Stalprodukt S.A ul. Wygoda 69, Bochnia. Polen. Dawid Barnaś
Statens vegvesen Stalprodukt S.A ul. Wygoda 69, 32-700 Bochnia Polen Dawid Barnaś Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet Matteo Pezzucchi /
DetaljerEn praktisk anvendelse av ITIL rammeverket
NIRF 17. april 2012 En praktisk anvendelse av ITIL rammeverket Haakon Faanes, CIA,CISA, CISM Internrevisjonen NAV NAVs ITIL-tilnærming - SMILI NAV, 18.04.2012 Side 2 Styring av tjenestenivå Prosessen omfatter
DetaljerUNIT LOG (For local use)
(EUROpean Pain Audit In Neonates) European survey of sedation and analgesia practices for ventilated newborn infants UNIT LOG (For local use) MONITORING OF INCLUSIONS/ EXCLUSIONS Principal Investigators
DetaljerStatens vegvesen. Godkjenning av LT 103 vegrekkverk. Linetech GmbH & Co. KG Von-Hünefeld-Straße Köln Tyskland
Statens vegvesen Linetech GmbH & Co. KG Von-Hünefeld-Straße 99 50829 Köln Tyskland Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet Matteo Pezzucchi /
DetaljerTEKNISK RAPPORT FRANZEFOSS MILJØKALK AS UTSLIPPSMÅLINGER - 2010 RAPPORT NR. 2010-3494 REVISJON NR. 01 DET NORSKE VERITAS
FRANZEFOSS MILJØKALK AS UTSLIPPSMÅLINGER - 2010 RAPPORT NR. 2010-3494 REVISJON NR. 01 DET NORSKE VERITAS Dato for første utgivelse: Prosjekt nr.: 2010-10-25 BA000818 Godkjent av: Organisasjonsenhet: Einar
DetaljerHan Ola of Han Per: A Norwegian-American Comic Strip/En Norsk-amerikansk tegneserie (Skrifter. Serie B, LXIX)
Han Ola of Han Per: A Norwegian-American Comic Strip/En Norsk-amerikansk tegneserie (Skrifter. Serie B, LXIX) Peter J. Rosendahl Click here if your download doesn"t start automatically Han Ola of Han Per:
DetaljerMCP-16RC, luftrenertårn
Kompakt filterkassett med pulsrengjøring MCP-16RC luftrensertårn er en kompakt filterkassett for desentralisert inneluftrengjøring der luftgjenoppretting er mulig. Den kompakte filterenheten leveres med
DetaljerStatens vegvesen. Fra VSI test senter: testrapportnummer KIR14003 datert TB32 test av
Statens vegvesen DELTA BLOC International GmbH Industriestrasse 28 2601 Sollenau Østerrike Paul Bittner Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet
DetaljerData Sheet for Joysticks
Available with Potentiometers or Hall sensors Several handle options Small size at low installation depth The 812 series is available with several different handle options. These small joysticks are recommended
DetaljerKurskategori 2: Læring og undervisning i et IKT-miljø. vår
Kurskategori 2: Læring og undervisning i et IKT-miljø vår Kurs i denne kategorien skal gi pedagogisk og didaktisk kompetanse for å arbeide kritisk og konstruktivt med IKT-baserte, spesielt nettbaserte,
DetaljerEFFEKTIV BRUK AV VIDEO I TRENING OG FORSKNING
Canon XL-1s med en Canon EF 100-400 mm objektiv Robert C. Reid, PhD Norges skiforbund hastighet bruk av video Før Filming: Hva er du ut etter å belyse? Hvordan vil du bruke opptaket etterpå (bilde sekvens,
Detaljer2A September 23, 2005 SPECIAL SECTION TO IN BUSINESS LAS VEGAS
2A September 23, 2005 SPECIAL SECTION TO IN BUSINESS LAS VEGAS SPECIAL SECTION TO IN BUSINESS LAS VEGAS 3A September 23, 2005 SEE, PAGE 8A Businesses seek flexibility. It helps them compete in a fast-paced,
DetaljerVELKOMMEN INN I DITT NYE TV-UNIVERS. Foto: Jens Haugen / ANTI
VELKOMMEN INN I DITT NYE TV-UNIVERS Foto: Jens Haugen / ANTI I esken / In the box 1 Huawei Q22 tv-dekoder / Huawei Q22 STB 2 3 4 HDMI 2.0-kabel / HDMI 2.0 cable Nettverkskabel / Ethernet cable Strømforsyning
DetaljerStatens vegvesen. Rekkverket har en totalbredde på 0,54m og høyde på 0,90m (uten kantdrager).
Statens vegvesen Likelydende brev - se vedlagt liste Behandlende enhet: Saksbehandler/telefon: Vår referanse: Deres referanse: Vår dato: Vegdirektoratet Matteo Pezzucchi / 22073246 15/212577-15 07.12.2015
DetaljerMidnight BBQ Light USER MANUAL
Midnight BBQ Light USER MANUAL Instructions The Midnight BBQ Light uses 4 x LR44 / AG13 batteries, included in the package. Unscrew the bottom cover and insert the included batteries and align the battery
DetaljerRE ( ) Tidsrelé plugg multi V
Produktdatablad Karakteristikk RE88867305 (41 726 84) Tidsrelé plugg multi 24-240V Alternativer Elektrisk tilkobling Kontaktmateriale [In] merkestrøm [Us] matespenning Voltage range Kapslingsmateriale
DetaljerKalibrering av DAP-meter
Kalibrering av Per Otto Hetland, SSDL, Statens strålevern Gardermoen, 09.11.2007 Innhold Hva er et? Hvorfor kalibrere og kalibrere/verifisere røntgenutstyr? Hvordan kalibreres ved SSDL, Statens strålevern
DetaljerLEAN PLANNING I PROSJEKTBASERT INDUSTRI. NIMA SCM Gabriele Hofinger Jünge
LEAN PLANNING I PROSJEKTBASERT INDUSTRI NIMA SCM Gabriele Hofinger Jünge 26.11.2015 I LØPET AV DE NESTE 20 MINUTTENE SKAL JEG FORTELLE OM Livet som forsker Forskningsprosjektet VALUE TANK LEAN planning
DetaljerSTOREFRONTS. Typical Details SCREW RACE JOINERY. Center Glazed Series B1. Interior Glazing (Shown with FF400 subsill) April 1998
- SCREW RCE JOINERY NOTE: NP22 glazing gaskets are used on both sides of glass. Typical TYPICL ELEVTION / (2.) HC20 Optional Head nchor PS00 Optional Filler EC0 Subsill s End Closure Typical (0.) IS IS2
DetaljerOm Strålevernets roller
Om Strålevernets roller Hilde M. Olerud, Dr.ing. Seksjonssjef, Statens strålevern fører tilsyn med bruk av strålekilder og spaltbart materiale koordinerer beredskap mot atomulykker og radioaktivt nedfall
DetaljerA/0860000^ '&-- 1985= 13
STATENS INSTITUTT FOR STRÅLEHYGIENE A/0860000^ '&-- 1985= 13 SAMMENLIKNING AV DOSIMETRIEN VED DE 4 PERSONDOSIMETRILABORATORIENE I NORGE Tor Wøhni National institute of radiation hygiene Østerndalen 25
Detaljer